ISO DIS 17862

2013 Edition, June 18, 2013

Complete Document

SURFACE CHEMICAL ANALYSIS - SECONDARY ION MASS SPECTROMETRY - LINEARITY OF INTENSITY SCALE IN SINGLE ION COUNTING TIME-OF-FLIGHT MASS ANALYSERS



View Abstract
Product Details
Document History

Detail Summary

Superseded By: ISO 17862

Additional Comments:
NOW A PUBLISHED STD * SEE ISO 17862
Format
Details
Price (USD)
Print
In Stock
Call for Quote
Add to Cart

Product Details:


Description / Abstract:

This International Standard specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity non-linearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead time. The correction can increase the intensity range for 95% linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction equations have been shown to be valid. This International Standard may also be used to confirm the validity of instruments in which the dead time correction is already made but in which further increases may or may not be possible.